Fluke - 718Ex Intrinsically Safe Pressure Calibrator

Product Overview:

  • ATEX II 1G Ex ia IIC T4 Compliant.
  • Built-in pressure / vacuum hand pump, with fine adjust vernier and bleed valve.
  • 30, 100, and 300 psi ranges (2, 7, and 20 bar)
  • Pressure measurement to 0.05% of full span, using an internal pressure sensor.
  • Pressure measurement to 200 bar using any of eight intrinsically safe Fluke 700PEx Pressure Modules.
  • Wide range of selectable measurement units for pressure.
  • Current measurement with 0.02% accuracy and 0.001 mA resolution.
  • Pressure switch test function.
  • Min-Max Hold functions.
  • Compact size and weight.
  • Simple, push-button user interface.
  • Rugged and reliable, ideal for field use.

More

Pump Up the Pressure

The Fluke 718Ex Intrinsically Safe Pressure Calibrator is a powerful intrinsically safe, self contained pressure calibrator for usage in explosion endangered areas.

Fluke 718Ex 30G Intrinsically Safe Pressure Calibrator
30 psi

Includes:

  • Protective red holster
  • TL75 test leads
  • AC72 test clips
  • CD users manuals (English, French, German, Spanish, Italian, Dutch, Norwegian, Danish, Swedish, Finnish, Portuguese, Korean, Chinese, Japanese)
  • Fluke 718Ex CCD content control drawing
  • NIST- traceable calibration report and dat

Fluke 718Ex 100G Intrinsically Safe Pressure Calibrator
100 psi

Includes:

  • Protective red holster
  • TL75 test leads
  • AC72 test clips
  • CD users manuals (English, French, German, Spanish, Italian, Dutch, Norwegian, Danish, Swedish, Finnish, Portuguese, Korean, Chinese, Japanese)
  • Fluke 718Ex CCD content control drawing
  • NIST- traceable calibration report and data

Fluke 718Ex 300G Intrinsically Safe Pressure Calibrator
300 psi

Includes:

  • Protective red holster
  • TL75 test leads
  • AC72 test clips
  • CD users manuals (English, French, German, Spanish, Italian, Dutch, Norwegian, Danish, Swedish, Finnish, Portuguese, Korean, Chinese, Japanese)
  • Fluke 718Ex CCD content control drawing
  • NIST- traceable calibration report and dat
Pressure Measurement Accuracy
Fluke 718Ex 30G Range -830 mbar to 2 bar (internal sensor)
Over pressure: 3x FS
Resolution 0.0001 bar
Accuracy 0.05% FS
Functions Zero, Min, Max, Hold, Damp
Fluke 718Ex 100G Range -830 mbar to 2 bar (internal sensor)
Over pressure: 2x FS
Resolution 0.001 bar
Accuracy 0.05% FS
Functions Zero, Min, Max, Hold, Damp
Fluke 718Ex 300G Range -830 mbar to 20 bar (internal sensor)
Over pressure: 375 psi, 25 bar
Resolution 0.0001 bar
Accuracy 0.05% FS
Functions Zero, Min, Max, Hold, Damp
Fluke 718Ex series Range Via pressure module connector
29 pressure modules, 2.5 mbar to 700 bar
Over pressure per pressure module specs
Resolution Per pressure module specs
Accuracy To 0.025% of full span, per pressure module specs
Functions Zero, Min, Max, Hold, Damp
Source pressure, built in pump range -820 mbar to full scale
For use with non-corrosive gasses.
Supported Pressure Units psi, inH2O (4°C), inH2O (20°C), kPa, cmH2O (4°C), cmH2O (20°C), bar, mbar, kg/cm², mmHg, inHg
Media compatibility per pressure module specs
Technical Data
Measure accuracy Range 0-24 mA
Resolution 0.001 mA
Accuracy 0.02% rdg ±2 counts
Display LCD, 5 digit pressure and current simultaneous
Environmental Specifications
Operating temperature -10°C to 55°C
Storage temperature -40°C to 60°C
Humidity (without condensation) 95% (10°C to 30°C)
75% (30°C to 40°C)
45% (40°C to 50°C)
35% (50°C to 55°C)
Operating altitude 3,000 m max
Safety Specifications
Safety ATEX II 1G Ex ia IIC T4 compliant
CSA Certification I.S. Class I
Div 1 Groups A-D , T4
Mechanical and General Specifications
Size (L x W x H) 216 x 94 x 66 mm (with holster)
Weight Approx. 992 g (with holster)
Warranty Three years, one year for pump
Shock 1 meter drop test
Vibration Random, 2 G, 5-500 Hz
Battery type One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V
Battery life 4 to 20 hours, depending on functions used

More Product Information

  • High performance intrinsically safe loop calibrator for use in potentially explosive environments
  • Is certified in accordance with ATEX directive (Ex II 2 G Ex ia IIC T4) in Zones 1 and 2 and to operate in accordance with Factory Mutual N.I. Class 1, Division 2 areas Group A-D
  • Features large display and simple, quick-click interface for easy one-handed operation

Intrinsically safe flashlights from Fluke make working in hazardous zones a little less hazardous. Rugged and reliable. Safety certified. From the largest supplier of intrinsically safe measurement tools.
  • UL certified – CLASS I DIV 1&2 GRP ABCD; CLASS II DIV 1&2 GRP EFG
  • Protected from harsh environments – IP67 rated
  • Built with Fluke’s reputation for safety, dependability and ruggedness
  • Part of Fluke’s market leading line of intrinsically safe products

Intrinsically safe flashlights from Fluke make working in hazardous zones a little less hazardous. Rugged and reliable. Safety certified. From the largest supplier of intrinsically safe measurement tools.
  • UL certified – CLASS I DIV 1&2 GRP ABCD; CLASS II DIV 1&2 GRP EFG
  • Protected from harsh environments – IP67 rated
  • Built with Fluke’s reputation for safety, dependability and ruggedness
  • Part of Fluke’s market leading line of intrinsically safe products

Intrinsically safe flashlights from Fluke make working in hazardous zones a little less hazardous. Rugged and reliable. Safety certified. From the largest supplier of intrinsically safe measurement tools.
  • UL certified – CLASS I DIV 1&2 GRP ABCD; CLASS II DIV 1&2 GRP EFG
  • Protected from harsh environments – IP6X rated
  • Built with Fluke’s reputation for safety, dependability and ruggedness
  • Part of Fluke’s market leading line of intrinsically safe products

Intrinsically safe flashlights from Fluke make working in hazardous zones a little less hazardous. Rugged and reliable. Safety certified. From the largest supplier of intrinsically safe measurement tools.

  • UL certified – CLASS I DIV 1&2 GRP ABCD; CLASS II DIV 1&2 GRP EFG
  • Protected from harsh environments – IP67 rated
  • Built with Fluke’s reputation for safety, dependability and ruggedness
  • Part of Fluke’s market leading line of intrinsically safe products

  • A completely sealed, IP67 rated case
  • Withstands drops up to 10 ft or 3 m (with holster)
  • Dustproof per IEC 60529 IP6x
  • Waterproof per IEC 60529 IPx7
  • CAT III 1000 V, CAT IV 600 V
  • Reversible holster for added protection of the display when not in use

  • Intrinsically safe infrared and contact thermometer
  • Meets intrinsically safe certifications in Class I Div. 1 and Div. 2 or Zone 1 and 2 hazardous environments
  • Measures temperatures from -40°C to 800°C (-40°F to 1472°F)
  • Comes with a conductive case to safely carry the IR thermometer into hazardous areas
  • Offers a distance-to-spot ratio of 50:1

  • Measure pressure, 0.05% of full scale with internal sensor up to 5000 psi/345 bar sensor (5000G model)
  • Measure mA with 0.015% accuracy and 0.001 mA resolution, while sourcing 24 V loop power
  • Measure pressure to 10,000 psi/700 bar using one of 29 Fluke 700Pxx Pressure Modules
  • 1/8 NPT female pressure fitting
  • Compatible with non-corrosive Gases and liquids
  • Select eleven different engineering units
  • Zero, min, max, and hold functions
  • Calibrate P/I instruments with simultaneous pressure and current measurement
  • Pressure switch test function captures set, reset and Deadband values of a pressure switch
  • Error % calculation of mA measurements for quick pass-fail decisions
  • Long lasting battery life
  • Holster features test lead storage
  • Compatible with Toolpack magnetic hanging system
  • New switch test

  • Pressure Source and Milliamp Measurement to calibrate and maintain almost any pressure device
  • Integrated pump is easily cleaned when accidently exposed to fluids that reduces cost of ownership and repairs and enables servicing the pump in the field
  • Error % calculation for quicker pass/fail decision making in the field
  • Best in class mA accuracy of 0.015% for better measurement confidence and workload coverage
  • Min/Max/Hold to capture changing measurements
  • Switch test feature for quicker and more accurate pressure switch testing
  • New 1 and 300 psi pressure ranges mean fewer extra tools required
  • 1, 30, 100 & 300 psi ranges now available

Fluke Corporation

Since its founding in 1948, Fluke has helped define and grow a unique technology market, providing testing and troubleshooting capabilities that have grown to mission critical status in manufacturing and service industries. Every new manufacturing plant, office, hospital, or facility built today represents another potential customer for Fluke products.

From industrial electronic installation, maintenance and service, to precision measurement and quality control, Fluke tools help keep business and industry around the globe up and running. Typical customers and users include technicians, engineers, metrologists, medical-device manufacturers, and computer network professionals — people who stake their reputations on their tools, and use tools to help extend their personal power and abilities.

Fluke has achieved the number one or number two position in every market in which it competes. The Fluke brand has a reputation for portability, ruggedness, safety, ease of use and rigid standards of quality.

Contact Details

Fluke Corporation Headquarters

6920 Seaway Blvd., Everett, WA 98203

Mailing address: PO Box 9090, Everett, Washington 98206-9090

Phone: 1-425-347-6100

Technical Support Email: fluke-info@fluke.com


Thermal Imaging
IR Windows Thermal cameras